1.
China Medical Equipment ; (12): 11-14,15, 2017.
Article in Chinese | WPRIM | ID: wpr-606387

ABSTRACT

Objective:To establish the X-ray standard was used to calibrate Hp(3) dosimeter in order to satisfy the requirement of eye-lens dose equivalents for the radiological occupational staff.Methods:The conventional values (Hp(3)) on the reference point of the narrow beams of X-ray fields were obtained by means of the product of air kerma (Kair) and the conversion coefficients (hp(3,ɑ)) recommended by ICRP116 recommendation. And these researches can be used by standard ionization chamber dosimeter. Finally, provided the conventional true value Hp(3) at calibrate point.Results:The standards and calibration program of calibrations eye-lens dose equivalent dosimeter were established under the X-ray radiation field. The detection error between slab phantom recommended by ISO and head phantom recommended by the 116th publication of ICRP was lower when the TLD was used to detect result.Conclusions:The standards for calibration Hp(3) dosimeters can satisfy the requirement of calibration for radiological occupational staff. It also provides theoretical foundation for detection data that assessed Hp(3) dosimeters and then improves trace ability and reliability of detection data of Hp(3) in national radiation supervision and inspection network.

2.
China Medical Equipment ; (12): 10-12,13, 2016.
Article in Chinese | WPRIM | ID: wpr-603931

ABSTRACT

Objective:According to the technical standard of ISO and current regulations in China, the X/γ standard calibration quality for the finger-ring dosimeter irradiation was established in Beijing SSDL, it can ensure the traceability of surveillance data within countrywide.Methods: Air kerma-rate was determined in the X-ray beams,137Cs and60Co by a standard dosimeter, combined transfer coefficientHp(0.07) provided by standards, and confirmed condition of X/γ standard calibration quality for the finger-ring dosimeter irradiation. Using finger-ring TLD dosimeter verified the radiation beam qualities Hp on the ISO finger-phantom.Results: The linearity and energy response of the TLD dosimeter met ISO and national standard in Beijing SSDL.Conclusion: The X/γ standard radiation field can be used for the calibration of finger-ring dosimeter standard curves and experiment of research, and can further promote the comparability and reliability of the surveillance data.

3.
China Medical Equipment ; (12): 4-6,7, 2015.
Article in Chinese | WPRIM | ID: wpr-601936

ABSTRACT

Objective:To complete experiments of the energy response compensation for three semiconductor personal dosimeters in order to meet the requirements of IEC and National Standard.Methods: In Beijing SSDL, the semiconductor detectors were covered by the different thicknesses of Pb and Sn additional filters with several diameter holes to compensate the dosimeters’ energy response.Results: At the 50 keV-Cs-137 energy range, the measuring error of dosimeters is met within 30% of energy response.Conclusion:By the compensation of combined Pb and Sn filters with a hole, the semiconductor personal dosimeters can be used for radiation monitoring.

4.
China Medical Equipment ; (12): 8-10,11, 2014.
Article in Chinese | WPRIM | ID: wpr-598796

ABSTRACT

Objective:According to the requirement of ISO and National Standard, new X-ray environmental standard for the calibration of environmental dosimeter was established in Beijing SSDL. Methods:Half value layer and air kerma dose-rate was determined in the X-ray environmental qualities by a standard dosimeter. Results: The technical requirement of ISO and National Standard are met within 2% for the X-ray environmental standard qualities in Beijing SSDL. Conclusion:The X-ray standard radiation field can be used for the calibration of environmental dosimeters and experiment of research.